1. ResMed Asia Product Support
  2. Technical Service
  3. Troubleshooting on AirSense 10, AirStart, AirCurve, Lumis

The following message is displayed on the device: "System fault, refer to user guide, Error 017 Description: EEPROM data error"

AirSense 10, AirStart, AirCurve, Lumis

Possible Cause:

  • Faulty EEPROM 
  • Software fault
Solution:
  • Re-install device software
  • If the above solution does not work, replace main PCB