The following message is displayed on the device: "System fault, refer to user guide, Error 004 Description: Main PCB over temperature"
AirSense 10, AirStart, AirCurve, Lumis
Possible Causes:
- Issue with device conditions
- Blocked filter
- Air tubing blocked
- Water in the air tubing
- Faulty main PCB
Solutions:
- Ensure ambient temperature conditions within specified operating range
- Ensure device is not covered.
- Allow device to cool before re-use. Disconnect power supply then reconnect to restart device
- Replace filter
- Allow device to cool down
- Power-cycle to remove message
- Check air tubing and remove any blockages. Press dial to clear message
- Empty water from air tubing. Disconnect power supply then reconnect to restart device.
- Replace main PCB if the above solutions do not work