SmartStart is enabled, but the device does not automatically start when the patient breathes into their mask
AirSense 10, AirStart, AirCurve, Lumis
Possible Cause:- Breath is not deep enough to trigger SmartStart
- There is excessive leak
Solution:
- To start therapy, take a deep breath in and out through the mask, before breathing normally
- Adjust the mask and headgear
- Air tubing may not be connected properly. Connect firmly at both ends