What if the patient feels they are not getting enough air?
AirMini / ActiveAir
Patients may feel they do not receive enough air when therapy starts as the AirMini has a default auto ramp setting starting at 5cmH2O. Auto ramp remains at 5cmH2O until AirMini detects sleep onset or times out (30 minutes) at which point it begins ramping up to full therapy pressure. For patients not satisfied with the default auto ramp setting, you should check the ramp settings of their other CPAP machine (if they have one) and match those on the AirMini.